Monitor Setup
2000966-386D Dashâ„¢ 3000/4000/5000 3-11
NOTE
Discharge patients from the monitor before restoring factory defaults.
1. Select MORE MENUS > MONITOR SETUP > MONITOR DEFAULTS.
2. Select CUSTOM DEFAULTS > RESTORE FACTORY DEFAULTS.
3. Position the pointer in front of the desired custom name.
4. Press the Trim Knob control.
Defining Parameter Window Priority
The number of parameter windows that display on the screen depends on the display
mode. No more than six full-sized parameter windows can display on the right side of
the monitor. No more than 10 automatically sized parameter windows can display in
total; six on the right side and four on the bottom of the screen.
NOTE
The large CLOCK DISPLAY option uses the space of one parameter window. If
the maximum number of parameter windows are needed for patient monitoring,
turn this option off.
NOTE
O
nly available in software version 6 or later.
You can define which parameters display as full-sized parameter windows in the
DEFAULT PARAMETER PRIORITY window by selecting the parameters in
priority order in the PARAMETER 2 to PARAMETER 6 fields.
579C
NOTE
PARAMETER 1 cannot be changed. It is always ECG. If PARAMETER 2 is set
for ECG or ST, the ECG parameter window will display the size of two
parameter windows.
You can also define which parameters display on the bottom of the screen by selecting
the parameters in the order that you want them to display from the list of parameters
on the bottom half of the DEFAULT PARAMETER PRIORITY window. The