Monitoring Temperature
2000966-386D Dashℱ 3000/4000/5000 14-3
NOTE
All changes are temporary and revert to the factory or custom defaults when the
patient is discharged. To permanently change the these settings, refer to Defining
Custom Defaults on page 3-7.
Troubleshooting
Option Function
T1 Select the T1 temperature site to acquire patient data. When
off, no patient data is acquired or displayed.
T2 Select the T2 temperature site to acquire patient data. When
off, no patient data is acquired or displayed.
UNITS Select the unit of measurement.
TEMPERATURE LIMITS Select the high and low alarm limits for each temperature
site. An alarm sounds when a value occurs outside of these
limits.
Message Cause Solution
CAL CHECK The monitor detects a 0.1ÂșC deviation
between the temperature value and the
internal calibration.
Check the sensor.
CAL FAIL
SENSOR The monitor does not detect a sensor.