A-10 Dash™ 3000/4000/5000 2000966-386D
Maintenance
Battery Charger
When you select the Target Capacity switch on the charger, the charger compares
the battery’s performance to a 60, 70, or 80 percent target capacity set on the battery
charger. If the battery fails to meet the target performance, the battery charger will
prompt you to condition the battery. If after the condition cycle has been completed,
the battery does NOT meet the target performance, then the battery charger will
illuminate a “fail” light.
Battery Status Menu
To display the BATTERY STATUS menu, select MORE MENUS > BATTERY
STATUS.
637B
Select BATTERY HELP for more information on batteries.
SLOT STATUS — Definitions of the information presented in the BATTERY
STATUS information window.
TIMES — Definitions of the information presented in the BATTERY STATUS
information window.
FUEL GAUGES — Explains the battery capacity gauges.
ALARMS — Lists the battery messages and solutions.
Installation Guidelines
Position the monitor in a location that does not artificially increase the operating
temperature of the battery.
To optimize battery life and performance, select a location that does NOT
artificially increase the ambient temperature surrounding the monitor.
Do not place the monitor near a heat vent or near heat-generating equipment,
such as computer monitors.

Table of Contents

Other manuals for GE Dash 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 3000 and is the answer not in the manual?

GE Dash 3000 Specifications

General IconGeneral
BrandGE
ModelDash 3000
CategoryMonitor
LanguageEnglish

Related product manuals