E-12 Dashâ„¢ 3000/4000/5000 2000966-386D
Interfacing with Non-GE Devices
Limitations
Duplicate Devices
The connectivity device can support a pre-defined number of each device type and up
to eight total devices, including the monitor. When the maximum number is exceeded
on any device type, the connectivity device ignores all data from the duplicate device.
îš„ One ventilator, anesthesia machine or gas analyzer.
îš„ Seven IV pumps.
Puritan-Bennett 840
Ventilator
BASELINE-000-R4-1
4-070212-85-F, -G, -
H, -K (English)
4-070000-85-L, -M, -P
(English)
4-070201-85-K, -L, -M
(Spanish)
4-070202-85-M, -N, -
P (French)
4-070203-85-L, -M, -N
(German)
4-070204-85-L, -M, -N
(Italian)
— VNT Yes, with
the
following
exceptions:
V
E TOT
V
TE MAND
V
TE SPONT
V
TI SPONT
420915-063 418265-063
Siemens Servo i
Ventilator
System software
version V3.01.02
VNT Pres,
CO
2
Exp
VNT Yes 420915-077 418265-077
GE Datex-Ohmeda
Ventilator
Engstrom software
version 4.15
VNT Pres,
VNT Flow,
CO
2
Exp
VNT, CO
2
Yes 420915-083 418265-083
Respironics Esprit
Ventilator
Esprit: 6.10, 7.10,
8.10
—VNT Yes
420915-082
418265-082
Stephan Christina
Ventilator
V1.9.4 — VNT Yes 420915-078 418265-078
LTV 950/1000 5.01 VNT Pres,
VNT Flow
VNT Yes 420915-086 418265-086
SLE 5000 3.0 — VNT Yes 420915-084 418265-084
GE Datex-Ohmeda
Aestiva/5, Aespire 7900
4.8 VNT Pres VNT Yes 420915-091 418265-091
GE Datex-Ohmeda
Centiva 5
V2.24.0, V3.00.0,
Com 1.0
VNT Pres2
VNT Flow
VNT Yes 420915-089 418265-089
Supported Ventilators
Supported Device
Supported Device
Software Version(s)
Waveform Parameter Alarms
Part Number for
Interface
Adapter
Part Number for
Interface
Installation
Instructions