Monitoring SpO2
2000966-386D Dashā„¢ 3000/4000/5000 11-9
Ohmeda and Masimo
The following settings control pulse oximetry monitoring for the Ohmeda and
Masimo configuration options.
Nellcor
The following settings control pulse oximetry monitoring for the Nellcor
configuration option.
RATE VOL Select the volume of the pulse rate tone to sound when an
SpO
2
pulse is detected. This is a variable pitch tone. The
tone pitch decreases as the patient’s saturation level
decreases. The lower the setting, the quieter the tone.
NOTE
Turning the pulse tone volume off also automatically
turns off the ECG QRS volume.
SPO2 LIMITS Select the high and low alarm limits for SpO
2
percentage
and pulse rate (beats per minute). An alarm sounds when a
value occurs outside of these limits.
SPEED Select a sweep speed for the displayed waveform.
Option Function
Option Function
PERSISTENT Turn the feature on or off. When on, a ā€œPROBE IS OFF THE
PATIENTā€ system status WARNING displays in the SpO
2
parameter window when the patient cable is disconnected.
The parameter window and waveform still display.

Table of Contents

Other manuals for GE Dash 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 3000 and is the answer not in the manual?

GE Dash 3000 Specifications

General IconGeneral
BrandGE
ModelDash 3000
CategoryMonitor
LanguageEnglish

Related product manuals