A-8 Dash™ 3000/4000/5000 2000966-386D
Maintenance
As the battery ages, the full charge capacity of the battery will degrade and be
permanently lost. As a result, the amount of charge that is stored and available for
use is reduced.
Cadex SMart Two+ Battery Charger
NOTE
To extend the life of batteries, GE recommends using the external Cadex SMart
Two+ battery charger to perform all battery care functions.
To use the external battery charger, the monitor must be running software version 2A
or later and the charger must be running software version 1.1 or later.
The following LEDs appear on the battery charger. The combination of LEDs and
their meaning change by illumination type (solid or flashing).
Safety
WARNING
PHYSICAL INJURY—Do NOT install the monitor above a patient.
Make sure the batteries are completely inserted and battery doors are
completely closed. Falling batteries could seriously or fatally injure
neonatal or other vulnerable patients.
WARNING
EXPLOSION OR FIRE —Using non-recommended batteries could
result is injury/burns to the patients or users. Only use batteries
recommended or manufactured by GE. The warranty can be voided
if non-recommended batteries are used.
LED Illuminated Flashing
RUN Charging in progress. Initializing the battery.
RUN and CONDITION Conditioning in progress.
READY Charging is complete.
READY and CONDITION Conditioning is complete
(pass target).
Conditioning is required.
FAIL Battery fault. Charger fault.
FAIL and CONDITION Conditioning is complete
(fail target).