GEDRAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
Chapter 4 - Functional Checks 4-12
4-3-2-5 Swivel/Brake Lock Caster
NOTE: This procedure applies to casters other than RIGHT REAR Caster.
Figure 4-10 Total Lock Casters
1) Press ‘a’. (see: Figure 4-9)
2) Rotate wheel until it locks (several locations where caster engages).
3) Verify Swivel Lock engages and Brake is ON.
4) Press ‘b’.(see: Figure 4-9).
5) Verify wheel is free to move.
CAUTION
!! CAUTION:
System weighs approximately 90kg. Pay attention for system stability when working with the
casters.