GEDRAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
Chapter 4 - Functional Checks 4-17
4-3-3-4 Power Off / Shutdown Check
1) When you shut down the system, enter the scan screen and lightly press the Power On/Off switch
(see: Figure 4-14) once. The System-Exit window is displayed.
2) The SYSTEM - EXIT menu, used when power off the unit, gives you these choices:
• Logoff
Use this button to log off the current user. The system remains ON and ready for a new user to log
on. If the Logoff button is dimmed, it indicates that no user is logged on to the unit at the moment.
• Shutdown
Use this button to shut down the system. The entire system will shut down. It is recommended to
perform a full shutdown at least once a week.
If the Shutdown button is dimmed, press the power button or Alt F10 to shut down the unit.
•Cancel
Use this button to exit from the System-Exit menu and return to the previous operation.
• Exit (Only available when logged in as GE Service with Service Dongle)
Select this button when you want to exit to the Windows Desktop.