GERAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
10-28 Section 10-8 - Electrical Safety Tests
10-8-6-2 Generic Procedure
The testing meter is connected from accessible protectively-earthed parts of the equipment to ground.
Measurements should be made with power line polarity normal and reversed, and with the neutral open
and closed. Record the highest reading of current.
NOTE: Color of Power outlet cable
i) USA and Canada
Hot: Black
Neutral: White
Ground: Green or Green-yellow
ii) Others
Hot: Brown
Neutral: Blue
Ground: Green-yellow
When using Dale 600/601 or a similar test instrument, its power plug may be inserted into the wall outlet
and the equipment under test is plugged into the receptacle on the panel of the meter. This places the
meter in the grounding conductor and the current flowing through protective earth will be indicated in
any of the current ranges. The maximum allowable limit for earth leakage current is shown in Table 10-
12.
Figure 10-19 Set Up for Test of Earth Leakage Current, UL60601-1/IEC 60601-1 Clause 19
Table 10-17 Typical Data Sheet for Earth Leakage Current
Tester Neutral Tester Polarity Switch Measured Earth leakage current
Enter Name of tested peripheral here:
Closed Normal
Closed Reversed
Open Normal
Open Reversed

Table of Contents

Other manuals for GE LOGIQ P9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P9 and is the answer not in the manual?

GE LOGIQ P9 Specifications

General IconGeneral
BrandGE
ModelLOGIQ P9
CategoryMedical Equipment
LanguageEnglish

Related product manuals