GEDRAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
Chapter 10 - Care & Maintenance 10-9
10-5-4-2 Air filter cleaning
Clean the system's air filters to ensure that a clogged filter does not cause the system to overheat and
reduce system performance and reliability. It is recommended the filters be cleaned every two weeks,
but the requirements will vary with environment.
Plastic Air filter
1) Pull the front cover with hand and pull out the front air filter.
Figure 10-1 Front Air filter location
2) Pull out the Bottom air filter.
Figure 10-2 Bottom Air filter location
3) Dust the filter with a vacuum cleaner and/or wash it with a mild soapy solution. If washed, rinse
and dry the filter before re-installation.
4) Put back the air filters and the front cover.
NOTE: For your convenience or if the air filter is too dirty, replacement filters are available. Refer to
Chapter 9 for the air filter replacement part number.
CAUTION
Be sure to lock the wheels before cleaning the air filters to avoid injury by any unexpected
movement of the system.
DO NOT operate the unit without the air filters in place.
Allow the air filters to dry thoroughly before re-installing them on the unit.

Table of Contents

Other manuals for GE LOGIQ P9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P9 and is the answer not in the manual?

GE LOGIQ P9 Specifications

General IconGeneral
BrandGE
ModelLOGIQ P9
CategoryMedical Equipment
LanguageEnglish

Related product manuals