GEDRAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
Chapter 3 - Setup Instructions 3-15
3-6-3 Connecting the Secondary “Patient” LCD Monitor
1) Power OFF/Shutdown the system as described in: Section 3-7-3 on page 3-25.
2) Connect the Secondary Monitor according to connection scheme (see: Figure 3-7 on page 3-16).
CAUTION
!! CAUTION:
Secondary LCD “Patient” Monitor MUST NEVER be connected to the LOGIQ P9/P7 ultrasound
systems mains supply directly!
Always connect it to the supplied Isolation Transformer!
The Secondary Monitor is the only item to be connected to the Transformer.
NOTICE
!! NOTICE:
Secondary Monitor is NOT intended for diagnostic use.
It is an additional device used to allow the patient to watch the proceedings.
Take your time to think about the best position of the monitor in your facilities.
Patients should be able to view the monitor easily and without having to bend or turn around.
NOTICE
!! NOTICE:
Connection to Secondary Monitor is via HDMI cable, but HDMI connector output does NOT carry audio
signal.

Table of Contents

Other manuals for GE LOGIQ P9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P9 and is the answer not in the manual?

GE LOGIQ P9 Specifications

General IconGeneral
BrandGE
ModelLOGIQ P9
CategoryMedical Equipment
LanguageEnglish

Related product manuals