D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
8-180 Section 8-46 - Replacement of battery pack for LP7-P9 R2 Battery option
Section 8-46
Replacement of battery pack for LP7-P9 R2 Battery option
8-46-1 Manpower
1 person, 30 minutes
8-46-2 Tools
Standard Phillips Screwdriver, Hex-Wrench, Socket driver
8-46-3 Removal Procedure
1) Unscrew 2 screws to remove POWER CORD FIX BRKT. Then disconnect POWER CORD and
remove REAR COVER.
2) Unscrew 5 screws to remove REAR CABLE COVER BRKT. Then disconnect all cables.
Figure 8-216 Removing POWER CORD FIX BRKT, POWER CORD and REAR COVER
Figure 8-217 Unscrewing 5 screws to remove REAR CABLE COVER BRKT and disconnect cables