4 - 44 Section 4-3 - Functional checks
4-3-7 Probe/Connectors Check
CAUTION
TAKE THE FOLLOWING PRECAUTIONS WITH THE PROBE CABLES:
- KEEP AWAY FROM THE WHEELS
- DO NOT BEND
- DO NOT CROSS CABLES BETWEEN PROBES
Table 4-5 Probe and Connectors Checks
Step Task Expected Result(s)
1. Press Probe on the Operator Panel. A list of the connected probes will pop up on the screen.
2. If not already selected. Use the trackball to select the
desired probe.
An application menu for the desired probe is listed on
the screen.
3. Trackball to the desired application
Press Select to launch the application.
To change application without changing the current
probe, press Appl. on the Operator Panel.
The selected application starts.
4. Verify no missing channels All channels are functioning.
5. Verify there's no EMI/RFI or artifacts specific to the
probe.
No EMI/RFI or artifacts.
6. Test the probe in each active connector slot.,
see 3-7-6 "Connecting Probes" on page 3-24
It will display pictorial data each time
7. Do a leakage test on the probe, see Section 10-6
"Electrical Safety Tests" on page 10-20
It passes the test.
8. Repeat this procedure for all available probes.

Table of Contents

Other manuals for GE Vivid E9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid E9 and is the answer not in the manual?

GE Vivid E9 Specifications

General IconGeneral
BrandGE
ModelVivid E9
CategoryMedical Equipment
LanguageEnglish

Related product manuals