Chapter 8 - Replacement procedures 8 - 121
8-6-14 Options Holder / Left or Right Support replacement
The Options Holder (the Left Support or Right Support) is the base piece that attaches to the upper
console to support either a Transvaginal Probe Holder or a Storage Tray. The VIVID E9 can support
one Left Options Holder and one Right Options Holder at the same time.
8-6-14-1 Preparations
When preparing for the replacement, you must perform the following steps:
1.) Power down the VIVID E9.
2.) Disconnect the Mains Power Cable from the wall outlet.
3.) Separate and remove the Storage Tray.
WARNINGWARNING
ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9.
WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO
VOLTAGE GREATER THAN 30 VOLTS:
1. FOLLOW LOCK OUT/TAG OUT PROCEDURES.
2. TURN OFF THE BREAKER.
3. UNPLUG THE VIVID E9.
4. MAINTAIN CONTROL OF THE POWER PLUG.
5. WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE, AS THERE ARE NO TEST POINTS TO
VERIFY ISOLATION. THE AMBER LIGHT ON THE OP PANEL ON/OFF BUTTON WILL TURN OFF.
BEWARE THAT THE MAIN POWER SUPPLY AND BACK END PROCESSOR MAY BE ENERGIZED EVEN IF THE POWER
IS TURNED OFF WHEN THE CORD IS STILL PLUGGED INTO THE AC OUTLET.
Signed
Date
TA G
&
LOCKOUT

Table of Contents

Other manuals for GE Vivid E9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid E9 and is the answer not in the manual?

GE Vivid E9 Specifications

General IconGeneral
BrandGE
ModelVivid E9
CategoryMedical Equipment
LanguageEnglish

Related product manuals