10 - 8 Section 10-5 - System maintenance
10-5-2-2 Peripheral/option checks
If any peripherals or options are not part of the VIVID E9 configuration, the check can be omitted.
Refer to the User Manual for a list of approved peripherals/options.
10-5-3 Input Power
10-5-3-1 Mains cable inspection
Table 10-4 GE Approved Peripheral/Hardware Option Functional Checks
Step Item Description
1. Media Verify media drive(s) read/write properlty. Clean if necessary.
2. B/W Printer Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
3. Color Printer Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
4. DICOM Verify that DICOM is functioning properly. Send an image to a DICOM device.
5. ECG Verify basic operation with customer
6. Footswitch Verify that the footswitch is functioning as programed.
Clean as necessary.
Table 10-5 Mains Cable Inspection
Step Item Description
1. Unplug Cord Disconnect the mains cable from the wall and the VIVID E9.
2. Inspect Inspect it and its connectors for damage of any kinds.
3.
Verify
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that
no strands may cause a short circuit.
4. Verify Inlet connector retainer is functional.