Chapter 4 - General procedures and Functional checks 4 - 47
4-3-10 Back End Processor checks
• If all the previous tests have been passed successfully, the Back End Processor is most likely OK.
4-3-11 Operator Panel Test
• The Operator Panel is tested when the VIVID E9 is powered up as part of the start-up scripts, run
at every start-up.
4-3-12 Peripheral checks
4-3-12-1 Printer checks
The internal printer is controlled from the P1 and P2 keys on the VIVID E9’s Operator Panel.
The factory default is:
• P1 for the internal printer
• P2 for external (network) printer
Table 4-7 Printer checks
Step Task Expected Result(s)
1. When scanning in 2D Color Mode, Press Freeze to
stop image acquisition.
Image scanning stops with the last picture on the
screen.
2. Press P1 on the Operator Panel The image displayed on the screen is printed on the
assigned printer.
3. Press P2 on the Operator Panel The image displayed on the screen is printed on the
assigned printer.
4. Check if the print quality on the pictures from both
printers are of expected quality.