9 Technical data
207723 rev 7
Gendex Dental Systems 123
9.7.3 Scatter radiation in ceph program
Ceph head was in ceph-PIO position (beam hitting the
secondary collimator, ear posts and CCD-sensor).
Radiation at technique factors 90kV, 13mA, 10sec was
measured just below the ear ports (equivalent to radiation
hitting the skull) to be 36mR.
Scatter radiation was measured during the scanning of the
ceph head. Full lateral ceph images were taken with 90kV,
13mA, 10sec scanning/exposure time and scatter radiation
was measured at 50 cm distance from the head.
The following table shows the scatter radiation levels
measured with the test instrument
Height Right ear Nose Left ear
central beam 0.025mR 0.034mR 0.066mR
15cm up from
central beam
0.021mR 0.022mR 0.059mR
15cm down
from central
beam
0.038mR 0.029mR 0.081mR
Approved: Laihonen Tuuli 2016-12-02 16:41
Reviewed: Nieminen Timo Antero 2016-12-02 15:31
Approved
See PDM system to determine the status of this document. Printed out: 2017-03-22 10:54:38
D507729, 7
Copyright © 2016 by PaloDEx Group Oy. All rights reserved.

Table of Contents

Other manuals for Gendex GXDP-700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gendex GXDP-700 and is the answer not in the manual?

Gendex GXDP-700 Specifications

General IconGeneral
BrandGendex
ModelGXDP-700
CategoryDental equipment
LanguageEnglish

Related product manuals