EasyManuals Logo

Gendex GXDP-700 User Manual

Gendex GXDP-700
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #37 background imageLoading...
Page #37 background image
3 Imaging programs
207723 rev 7
Gendex Dental Systems 29
You can shift the Sinus Program image layer 10 mm in a
posterior direction by pressing the Image Layer- Protrusion
button.
3.3 Default exposure settings
User Configurable Panoramic mA Level
Default mA level for panoramic programs can be set using
the touch screen display. To change the setting:
1. Start an exam and select a panoramic program.
2. Select Settings on the touch screen.
3. Select Imaging program defaults.
4. Select Set current PAN mA level as default.
The selected mA value is used as default value for the
current program. The default mA level is adjusted by an
equal amount also for other panoramic programs.
3.4 User Configurable Default Program
Default imaging program (pan/ceph/3D) can be set using
the touch screen display. To change default program:
1. Start an exam and select the desired program.
2. Press the Settings button on the touch screen.
3. Select Imaging program defaults.
4. Select Set current program as default.
The selected program is automatically activated when an
exam is started for a new patient and at startup.
Approved: Laihonen Tuuli 2016-12-02 16:41
Reviewed: Nieminen Timo Antero 2016-12-02 15:31
Approved
See PDM system to determine the status of this document. Printed out: 2017-03-22 10:54:38
D507729, 7
Copyright © 2016 by PaloDEx Group Oy. All rights reserved.

Table of Contents

Other manuals for Gendex GXDP-700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gendex GXDP-700 and is the answer not in the manual?

Gendex GXDP-700 Specifications

General IconGeneral
BrandGendex
ModelGXDP-700
CategoryDental equipment
LanguageEnglish

Related product manuals