3 Imaging programs
207723 rev 7
Gendex Dental Systems 29
You can shift the Sinus Program image layer 10 mm in a
posterior direction by pressing the Image Layer- Protrusion
button.
3.3 Default exposure settings
User Configurable Panoramic mA Level
Default mA level for panoramic programs can be set using
the touch screen display. To change the setting:
1. Start an exam and select a panoramic program.
2. Select Settings on the touch screen.
3. Select Imaging program defaults.
4. Select Set current PAN mA level as default.
The selected mA value is used as default value for the
current program. The default mA level is adjusted by an
equal amount also for other panoramic programs.
3.4 User Configurable Default Program
Default imaging program (pan/ceph/3D) can be set using
the touch screen display. To change default program:
1. Start an exam and select the desired program.
2. Press the Settings button on the touch screen.
3. Select Imaging program defaults.
4. Select Set current program as default.
The selected program is automatically activated when an
exam is started for a new patient and at startup.