19-104 Return to Section Topics 2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands Series 2600A System SourceMeter® Instruments Reference Manual
smuX.cal.date
X = SMU channel (a or b)
Attribute Calibration date for the active calibration set.
Default 0
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage caldate = smuX.cal.date -- Reads calibration date.
smuX.cal.date = caldate -- Writes calibration date.
caldate Sets the current calibration’s adjustment date.
Remarks smuX.cal.date must be set to the date the adjustment was done using the UTC time and
date. The date is stored as the number of seconds since UTC 12:00 am Jan 1, 1970.
This attribute stores the calibration date associated with the active calibration set. The calibration
date can be read at any time but can only be assigned a new value when calibration has been
enabled with the smuX.cal.unlock function.
This attribute is stored with the active calibration set. If a different calibration set is restored, this
attribute will reflect the date stored with that set.
Due to the internal storage format, smuX.cal.adjustdate is only accurate to within a few minutes
of the value set.
Details See Calibration in Section 20.
Also see smuX.cal.adjustdate, smuX.cal.due, smuX.cal.restore, smuX.cal.save
Example Sets calibration date for SMU A to the current time set on the instrument:
smua.cal.date = os.time()
smuX.cal.due
X = SMU channel (a or b)
Attribute Calibration due date for the next calibration.
Default 0
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage caldue = smuX.cal.due -- Reads calibration due date.
smuX.cal.due = caldue -- Writes calibration due date.
caldue Sets the next calibration due date.
Remarks smuX.cal.due must be set to the date the adjustment was done using the UTC time and date.
The date is stored as the number of seconds since UTC 12:00 am Jan 1, 1970.
This attribute stores the calibration due date associated with the active calibration set. The
calibration due date can be read at any time but can only be assigned a new value when
calibration has been enabled with the smuX.cal.unlock function.
This attribute is stored with the active calibration set. If a different calibration set is restored, this
attribute will reflect the due date stored with that set.
Due to the internal storage format, smuX.cal.adjustdate is only accurate to within a few minutes
of the value set.
Details See Calibration in Section 20.
Also see smuX.cal.date, smuX.cal.restore, smuX.cal.state
Example
Sets calibration due date for one year from the current time set on the instrument:
smua.cal.due = os.time() + 365 * 24 * 60 * 60

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the Keithley 2600a series and is the answer not in the manual?

Keithley 2600a series Specifications

General IconGeneral
CategorySource Measure Unit (SMU)
Model2600A Series
TypePrecision SourceMeter
Voltage Range±200V
Current Range±10fA to ±10A (depending on model)
DC Current±10A
Pulse Current±100A
Maximum Power20W to 200W (depending on model)
Power per Channel20W
Accuracy0.012% basic
InterfaceGPIB, USB, Ethernet
Power Supply100V to 240V, 50Hz to 60Hz
Weight (1-channel)4.5kg
Channels1 or 2
Dimensions (1-channel)89 mm high x 214 mm wide x 429 mm deep (3.5 in x 8.4 in x 16.9 in)
Dimensions (2-channel)89 mm high x 214 mm wide x 429 mm deep (3.5 in x 8.4 in x 16.9 in)