19-230 Return to Section Topics 2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands Series 2600A System SourceMeter® Instruments Reference Manual
tspnet.tsp.abortonconnect
Attribute Abort on connect.
Default 1
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage tspnet.tsp.abortonconnect = value
value = tspnet.tsp.abortonconnect
value 1 or 0
Remarks • This setting determines if the Series 2600A sends an abort message when it attempts to connect
using tspnet.connect to a TSP
TM
-enabled instrument.
• Sending the abort command on a connection causes any other active interfaces being used on
that instrument to close to ensure you have obtained access to the remote instrument.
• Connecting to a TSP instrument without issuing an abort command, or when
tspnet.tsp.abortonconnect is set to 0, can result in the Series 2600A suspending
operation until it receives a response back from the instrument or until a timeout error is
generated.
Example tspnet.tsp.abortonconnect = 0

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2600a series and is the answer not in the manual?

Keithley 2600a series Specifications

General IconGeneral
BrandKeithley
Model2600a series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals