System SourceMeter® Instrument Reference Manual Section 7:
2600BS-901-01 Rev. B / May 2013 7-35
ConfigPulseIMeasureVSweepLin()
This KIPulse factory script function configures a linear pulsed current sweep with a voltage measurement at each
point.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
f, msg = ConfigPulseIMeasureVSweepLin(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in, sync_out, sync_in_timeout, sync_in_abort)
f, msg = ConfigPulseIMeasureVSweepLin(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in, sync_out, sync_in_timeout)
f, msg = ConfigPulseIMeasureVSweepLin(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in, sync_out)
f, msg = ConfigPulseIMeasureVSweepLin(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in)
f, msg = ConfigPulseIMeasureVSweepLin(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag)
A Boolean flag; this flag is true if the pulse was successfully configured, false when errors
were encountered
A string message; if the f flag is false, msg contains an error message; if it is true, msg
contains a string indicating successful configuration
System SourceMeter
®
instrument channel (for example, smua refers to SMU channel A)
Pulse sweep start level in amperes
Pulse sweep stop level in amperes
Voltage limit (for example, compliance) in volts
toff
Pulse off time in seconds
Number of pulse-measure cycles
Reading buffer where pulsed measurements will be stored; if this is nil when the function is
called, no measurements will be made when the pulse train is initiated
Numeric identifier to be assigned to the defined pulse train
Defines a digital I/O trigger input line; if programmed, the pulse train will wait for a trigger input
before executing each pulse
Defines a digital I/O trigger output line; if programmed, the pulse train will generate a trigger
output immediately before the start of
Specifies the length of time (in seconds) to wait for input trigger; default value is 10 s
Specifies whether or not to abort pulse if input trigger is not received; if pulse aborts because of
a missed trigger, a timer timeout message is returned;
or