Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-342 2600BS-901-01 Rev. B / May 2013
trigger.clear()
This function clears the command interface trigger event detector.
Type TSP-Link accessible
Affected by Where saved Default value
Function
No
Usage
trigger.clear()
Details
The trigger event detector indicates if a trigger event has been detected since the last trigger.wait() call.
trigger.clear() clears the trigger event detector and discards the history of command interface trigger
events.
Also see
trigger.wait() (on page 7-350)
trigger.EVENT_ID
This constant contains the command interface trigger event number.
Type TSP-Link accessible
Affected by Where saved Default value
Constant
Yes
Usage
eventID = trigger.EVENT_ID
eventID
The event ID for the command interface triggers
Details
You can set the stimulus of any trigger object to the value of this constant to have the trigger object respond to
command interface trigger events.
Example
trigger.timer[1].stimulus = trigger.EVENT_ID
Sets the trigger stimulus of trigger
timer 1 to the command interface
trigger event.
Also see
None

Table of Contents

Other manuals for Keithley 2611B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2611B and is the answer not in the manual?

Keithley 2611B Specifications

General IconGeneral
BrandKeithley
Model2611B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals