Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-208 2600BS-901-01 Rev. B / May 2013
Also see
Range (on page 2-82)
reset() (on page 7-160)
setup.recall() (on page 7-180)
smuX.measure.autorangeY (on page 7-198)
smuX.reset() (on page 7-216)
smuX.measure.nplc
This command sets the integration aperture for measurements.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes
Instrument reset
SMU reset
Recall setup
Saved setup 1.0
Usage
nplc = smuX.measure.nplc
smuX.measure.nplc = nplc
nplc
The integration aperture; set from 0.001 to 25
X
Source-measure unit (SMU) channel (for example, smua.measure.nplc applies
to SMU channel A)
Details
This attribute controls the integration aperture for the analog-to-digital converter (ADC).
The integration aperture is based on the number of power line cycles (NPLC), where 1 PLC for 60 Hz is
16.67 ms (1/60) and 1 PLC for 50 Hz is 20 ms (1/50).
Example
smua.measure.nplc = 0.5
Sets the integration time for SMU channel A to 0.5/60
seconds.
Also see
reset() (on page 7-160)
setup.recall() (on page 7-180)
smuX.reset() (on page 7-216)
Speed (on page 2-88)

Table of Contents

Other manuals for Keithley 2611B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2611B and is the answer not in the manual?

Keithley 2611B Specifications

General IconGeneral
BrandKeithley
Model2611B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals