Series 2600B
System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. B / May 2013 7-173
Details
The loadscript or loadandrunscript and endscript keywords are not included in the source code.
The body of the script is a single string with lines separated by the new line character.
The instrument automatically stores the source for all scripts that are loaded on the instrument. To free up
memory or to obfuscate the code, assign nil to the source attribute of the script. Although this attribute is
writable, it can only be set to the nil value.
Example
test7 = script.new("display.clear() display.settext('Hello from my test')", "")
print(test7.source)
This example creates a script called "test7" that displays a message on the front panel and retrieves the
source code.
Output:
display.clear() display.settext('Hello from my test')
Also see
scriptVar.list() (on page 7-169)
serial.baud
This attribute configures the baud rate for the RS-232 port.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Not applicable
Nonvolatile memory
9600
Usage
baud = serial.baud
serial.baud = baud
baud
The baud rate (300, 600, 1200, 2400, 4800, 9600, 19200, 38400, 57600 or
115200)
Details
A new baud rate setting takes effect when the command to change it is processed.
Allow ample time for the command to be processed before attempting to communicate with the
instrument again. If possible, set the baud rate from one of the other command interfaces or from the
front panel.
The reset function has no effect on data bits.
Example
serial.baud = 1200
Sets the baud rate to 1200.
Also see
RS-232 interface operation (on page 2-108)
serial.databits (on page 7-174)
serial.flowcontrol (on page 7-174)
serial.parity (on page 7-175)

Table of Contents

Other manuals for Keithley 2611B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2611B and is the answer not in the manual?

Keithley 2611B Specifications

General IconGeneral
BrandKeithley
Model2611B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals