EasyManuals Logo
Home>Keithley>Measuring Instruments>2611B

Keithley 2611B Reference Manual

Keithley 2611B
864 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #532 background imageLoading...
Page #532 background image
Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-192 2600BS-901-01 Rev. B / May 2013
Also see
Adjustment (on page B-18)
smuX.cal.restore() (on page 7-188)
smuX.cal.save() (on page 7-189)
smuX.cal.unlock() (on page 7-190)
smuX.contact.calibratelo() (on page 7-192)
smuX.contact.calibratelo()
This function calibrates the low/sense low contact check measurement. This function is not available on the
Models 2604B/2614B/2634B.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
smuX.contact.calibratelo(cp1Measured, cp1Reference, cp2Measured, cp2Reference)
X
Source-measure unit (SMU) channel (for example,
smua.contact.calibratelo()
applies to SMU channel A)
cp1Measured
The value measured by this SMU for calibration point 1
cp1Reference
The reference measurement for calibration point 1 as measured externally
cp2Measured
The value measured by this SMU for calibration point 2
cp2Reference
The reference measurement for calibration point 2 as measured externally
Details
Contact check measurement calibration does not require range information.
Typically, calibration points one and two are near 0 Ω and 50 Ω, respectively.
All four measurements (cp1Measured, cp1Reference, cp2Measured, and cp2Reference) must be made
with the active calibration set. If not, corruption of the calibration constants may result.
The new calibration constants are activated immediately but are not written to nonvolatile memory. Use
smuX.cal.save() to save the new constants to nonvolatile memory. The active calibration constants stay in
effect until the instrument is power cycled or a calibration set is loaded from nonvolatile memory with the
smuX.cal.restore() function.
This function is disabled until a successful call to smuX.cal.unlock() is made.

Table of Contents

Other manuals for Keithley 2611B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2611B and is the answer not in the manual?

Keithley 2611B Specifications

General IconGeneral
BrandKeithley
Model2611B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals