Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-212 2600BS-901-01 Rev. B / May 2013
smuX.measure.rel.levelY
This attribute sets the offset value for relative measurements.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes
Instrument reset
SMU reset
Recall setup
Not saved 0
Usage
relValue = smuX.measure.rel.levelY
smuX.measure.rel.levelY = relValue
relValue
Relative measurement offset value
X
Source-measure unit (SMU) channel (for example, smua.measure.rel.levelv
applies to SMU channel A)
Y
SMU measurement function (
v
= voltage,
i
= current,
r
= resistance,
p
= power)
Details
This attribute specifies the offset value used for relative measurements. When relative measurements are
enabled (see smuX.measure.rel.enableY), all subsequent measured readings will be offset by the value of
this attribute. Each returned measured relative reading is the result of the following calculation:
Relative reading = Actual measured reading - Relative offset value
Example
smua.measure.rel.levelv = smua.measure.v()
Performs a voltage measurement using
SMU channel A and then uses it as the
relative offset value.
Also see
Relative offset (on page 3-1)
reset() (on page 7-160)
smuX.measure.rel.enableY (on page 7-211)
smuX.reset() (on page 7-216)

Table of Contents

Other manuals for Keithley 2611B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2611B and is the answer not in the manual?

Keithley 2611B Specifications

General IconGeneral
BrandKeithley
Model2611B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals