Series 2600B
System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. B / May 2013 7-179
Example
settimezone("8", "1", "3.3.0/02", "11.2.0/02")
settimezone(offset)
Sets offset to equal +8 hours, +1
hour for DST, starts on Mar 14 at 2:00
a.m, ends on Nov 7 at 2:00 a.m.
Sets local time zone to offset.
Also see
gettimezone() (on page 7-98)
os.time() (on page 7-151)
settime() (on page 7-177)
setup.poweron
This attribute specifies which saved setup to recall when the instrument is turned on.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Not applicable
Nonvolatile memory
0
Usage
id = setup.poweron
setup.poweron = id
id
An integer that specifies the setup to recall when the instrument power is
turned on (0 to 5)
Details
When id = 0, the instrument uses the factory default setup when it is turned on. When id is set to 1 to 5, it uses
the setup saved with setup.save().
Only setups stored in nonvolatile memory are available (you cannot recall a script from a USB flash drive with
this command).
Example
setup.poweron = 0
Set the instrument to use the factory default setup
when power is turned on.
Also see
setup.save() (on page 7-181)
Start-up (power-on) configuration (on page 2-48)

Table of Contents

Other manuals for Keithley 2611B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2611B and is the answer not in the manual?

Keithley 2611B Specifications

General IconGeneral
BrandKeithley
Model2611B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals