Series 3700A System Switch/Multimeter Reference Manual Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018 11-41
Details
There is only one password per card. Therefore, channel.calibration.password() works only
on the currently unlocked card.
Make note of the password, because there is no command to query for the password once it has
been set on the instrument. The password is not permanently saved until the
channel.calibration.save() command is sent. Passwords are alphanumeric and case-
sensitive.
This command generates an error if calibration is locked or if the password string length is greater
than six characters.
The default password from the factory is KI3706.
Example
channel.calibration.unlock("slot3","KI3706")
channel.calibration.password("Unlock")
channel.calibration.save()
channel.calibration.lock()
Unlock the calibration for the card in slot 3,
assuming the default password is still valid.
Set the password to "Unlock" for the card in
slot 3.
Save the password for the card in slot 3 for
subsequent unlocks.
Lock the calibration for the card in slot 3.
Also see
channel.calibration.lock() (on page 11-40)
channel.calibration.unlock() (on page 11-44)
channel.calibration.save()
This function saves the calibration data to the card.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
channel.calibration.save()
Details
Only one card can be calibrated at a time. Therefore, channel.calibration.save() works only
on the presently unlocked card. An error is generated if this command is issued when calibration is
already locked.
The system must receive this command before the channel.calibration.lock() command or
the calibration data will be lost.
This command saves the present values of the calibration constants and calibration date, and
increases the calibration count by one, regardless of errors in the data. You should not issue
channel.calibration.save() unless the calibration procedure was performed with no errors.
If no calibration date was specified using either channel.calibration.adjustdate() or
channel.calibration.verifydate(), the date is automatically assigned based on the system
date.

Table of Contents

Other manuals for Keithley 3720-ST

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3720-ST and is the answer not in the manual?

Keithley 3720-ST Specifications

General IconGeneral
BrandKeithley
Model3720-ST
CategoryMultimeter
LanguageEnglish

Related product manuals