Section 11: TSP command reference Series 3700A System Switch/Multimeter Reference Manual
11-406 3700AS-901-01 Rev. D/June 2018
trigger.blender[N].clear()
This function clears the blender event detector and resets the overrun indicator of blender N.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
trigger.blender[N].clear()
N
The blender number (up to two)
Details
This command sets the blender event detector to the undetected state and resets the overrun
indicator of the event detector.
Example
trigger.blender[2].clear()
Clears the event detector for blender 2.
Also see
None
trigger.blender[N].EVENT_ID
This constant contains the trigger blender event number.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Constant
Yes
Usage
eventID = trigger.blender[N].EVENT_ID
eventID
Trigger event number
N
The blender number (up to two)
Details
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to
trigger events from this trigger blender.
Example
digio.trigger[1].stimulus = trigger.blender[2].EVENT_ID
Set the trigger stimulus of
digital I/O trigger 1 to be
controlled by the trigger
blender 2 event.
Also see
None

Table of Contents

Other manuals for Keithley 3720-ST

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3720-ST and is the answer not in the manual?

Keithley 3720-ST Specifications

General IconGeneral
BrandKeithley
Model3720-ST
CategoryMultimeter
LanguageEnglish

Related product manuals