Series 3700A System Switch/Multimeter Reference Manual Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018 11-239
eventlog.enable
This attribute enables or disables the event log.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Recall setup
Create configuration script
Save setup
eventlog.ENABLE
Usage
status = eventlog.enable
eventlog.enable = status
status
The enable status of the event log:
1 or eventlog.ENABLE: Event log enable
0 or eventlog.DISABLE: Event log disable
Details
When the event log is disabled (eventlog.DISABLE or 0), no new events are added to the event
log. You can, however, read and remove existing events.
When the event log is enabled, new events are logged.
Example
print(eventlog.enable)
eventlog.enable = eventlog.DISABLE
print(eventlog.enable)
Displays the present status of the Series
3700A event log.
Output:
1.00000e+00
0.00000e+00
Also see
eventlog.all() (on page 11-237)
eventlog.clear() (on page 11-238)
eventlog.count (on page 11-238)
eventlog.next() (on page 11-240)
eventlog.overwritemethod (on page 11-241)

Table of Contents

Other manuals for Keithley 3720-ST

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3720-ST and is the answer not in the manual?

Keithley 3720-ST Specifications

General IconGeneral
BrandKeithley
Model3720-ST
CategoryMultimeter
LanguageEnglish

Related product manuals