dmm.inputdivider
This attribute determines when the 10 MΩ input divider is enabled.
Instrument reset
DMM reset
Recall setup
Create configuration script
Save setup
Usage
state = dmm.inputdivider
dmm.inputdivider = state
Disable input divider: dmm.OFF or 0
Enable input divider: dmm.ON or 1
Details
This attribute is only valid when dmm.func is set to DC volts.
The input divider setting is saved with the dmm.func function setting, so if you use another function,
then return to "dcvolts", the input divider setting you set previously is retained.
An error is generated if you try to set input divider for any DMM function other than "dcvolts". Error
code 1114, "Setting conflicts with function selected," is generated. If you query any DMM function
other than "dcvolts" for input divider, nil is returned with the same error.
Example
dmm.func = "dcvolts"
dmm.inputdivider = dmm.ON
Enables the input divider for DC volts.