Series 3700A System Switch/Multimeter Reference Manual Section 4: Basic DMM operation
3700AS-901-01 Rev. D/June 2018 4-55
Range
You can use the range to set an expected measurement value. The instrument will select the range
appropriate to measure that value.
If you set a range, the autorange feature is automatically disabled. The instrument selects the range
to best match the expected measure value for the functions, as shown below.
The range setting is saved with the DMM function setting, so if you select another function, then
return to the previous function, the range settings you set previously are retained.
You cannot select a range that includes different channel types.
A power cycle or an instrument reset will clear the saved ranges.
Measurement ranges and maximum readings
The range that is selected affects both measurement accuracy and the maximum measurable level.
Input values that exceed the maximum readings cause an "Overflow" message to be displayed.
Function
Ranges
Maximum reading
DCV (DC voltage)
100 mV, 1 V, 10 V, 100 V, 300 V
±303 V
ACV (AC voltage)
100 mV, 1 V, 10 V, 100 V, 300 V
±303 V
DCI (DC current)
10 µA, 100 µA, 1 mA, 10 mA,
100 mA, 1 A, 3 A
±3.1 A
ACI (AC current)
1 mA, 10 mA, 100 mA, 1 A, 3 A
±3.1 A
2 (2-wire ohm)
10 Ω, 100 Ω, 1 kΩ, 10 kΩ,
100 kΩ, 1 MΩ, 10 MΩ, 100
120
4 (4-wire ohm)
1 Ω, 10 Ω, 100 Ω, 1 kΩ, 10 kΩ,
100 kΩ, 1 MΩ, 10 MΩ, 100
120
4 OC (4-wire ohm offset
compensated)
1 Ω, 10 Ω, 100 Ω, 1 , 10
12
4 DRY+ (4-wire ohm dry
circuit)
1Ω, 10 Ω, 100 Ω, 1 kΩ, 10
2.4
TMP (temperature)
200 °C to 1820 °C
Sensor dependent
FREQ (frequency)
100 mV, 1 V, 10 V, 100 V, 300 V
3 Hz to 500 kHz
PER (period)
100 mV, 1 V, 10 V, 100 V, 300 V
2 µs to 333 ms
CNT (continuity)
1
Threshold adjustable 1 Ω to
1000 Ω
CSΩ (commonside ohm)
1 Ω, 10 Ω, 100 Ω, 1 Ω, 10 kΩ,
100 kΩ, 1 MΩ, 10 MΩ, 100
120
Temperature
There is no range selection for temperature measurements, which are performed on a single fixed
range. Depending on the sensor, the maximum temperature readings range from 200 °C to 1820 °C.

Table of Contents

Other manuals for Keithley 3720-ST

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3720-ST and is the answer not in the manual?

Keithley 3720-ST Specifications

General IconGeneral
BrandKeithley
Model3720-ST
CategoryMultimeter
LanguageEnglish

Related product manuals