Probes and Biopsy 12-13
Immerse the intra-operative transducer in the solution (for reference)
CAUTION:
Repeated disinfection or sterilization will eventually damage
the transducer, please check the transducer’s performance
periodically.
Compatible Cleaner, Disinfectants and Sterilants
For the cleaner, disinfectants and sterilants information, please refer to Mindray Transducer
Disinfectant Recommendation.
12.1.6 Environment
Working condition
Use the probes in the following ambient conditions:
Probe
Ambient
temperature
Relative humidity
Atmospheric
pressure
35C50EA
0-40℃
30%-85%RH(no
condensation)
700-1060hPa
65C15EA
0℃-40℃
30%~85%RH(no
condensation)
700 hPa ~ 1060 hPa
65EC10EA
0℃-40℃
30%~85%RH(no
condensation)
700 hPa ~ 1060 hPa
65EB10EA
0℃~40℃
30%~90%RH
(no
condensation)
700hPa~1060hPa
75L38EA
0℃-40℃
30%~85%RH(no
condensation)
700 hPa ~ 1060 hPa
75L53EA
0℃~40℃
30%~85%RH
(no
condensation)
700hPa~1060hPa
10L24EA
0℃~40℃
30%~90%RH
(no
condensation)
700hPa~1060hPa
35C20EA
0℃-40℃
30%~85%RH(no
condensation)
700 hPa ~ 1060 hPa