EasyManuals Logo
Home>Mindray>Medical Equipment>DP-50

Mindray DP-50 User Manual

Mindray DP-50
245 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #218 background imageLoading...
Page #218 background image
13-2 Battery
NOTE:
1.
Only use the specified batteries.
2.
If there is only one battery in the system, it cannot supply power and
cannot be charged.
13.2 Precautions
1. Before using the battery, carefully read the description in the label on the surface of the
battery.
2. When you use the battery at the first time and find that it is dirty or emit an odor, do not
use it.
3. Do not expose the battery to heat sources or high pressure.
4. Do not leave the battery in direct sunlight.
5. Store the battery out of the reach of children. Store the battery out of the reach of children.
6. The battery is designed to be charged only in this system; charge the battery only when
the ambient temperature is between 0 and 40 .
7. When the battery is out of power, please charge it immediately.
8. For storage longer than three months, charge a battery to a level no more than 45%.
9. Do not knock the battery.
13.3 Installing and Removing the Batteries
WARNING:
Turn off the system and disconnect the adapter from the
main unit before installing or removing batteries. DO NOT
install or remove batteries when the system is turned on or
the system is turned off but the adapter is connected with
the system.
The battery is put in the battery bays at the back side of the system. You can install or remove
it when the system is turned off and the power cord is disconnected from the main unit.
To install the battery:
1. Turn off the unit and detach the power cord from the main unit.
2. You can see a battery bay at the back of the system. (Note that the battery shall be put in
correct direction). Put the battery into the bay and push to the right side until its tightly
locked.
3. Close the battery cover.
To remove the battery:
1. Turn off the unit and detach the power cord from the main unit.
2. Open the battery cover.
3. Push the battery to left until its released.
4. Take out the battery from the bay.
13.4 Battery Status Indicator
The battery status indicator is located in the lower right corner of the screen, indicating the
battery capacity.
: indicates the battery capacity is empty.