RIGOL 2 Command System
2-116 DS2000E Programming Guide
:FUNCtion:WANalyze:SETup:SFRame
:FUNCtion:WANalyze:SETup:EFRame
Syntax
:FUNCtion:WANalyze:SETup:SFRame <sframe>
:FUNCtion:WANalyze:SETup:SFRame?
:FUNCtion:WANalyze:SETup:EFRame <eframe>
:FUNCtion:WANalyze:SETup:EFRame?
Description
Set or query the start frame of waveform analysis.
Set or query the end frame of waveform analysis.
Parameter
Name Type Range Default
<sframe> Integer 1 to the number of frames recorded 1
<eframe> Integer 1 to the number of frames recorded The number of frames recorded
Explanation
The commands are only valid when the waveform recording is finished.
For the number of frames recorded, please refer to the
:FUNCtion:WREPlay:FMAX? command.
The start frame (end frame) of waveform analysis should be lower than or equal to (greater than or
equal to) the end frame (start frame) of waveform analysis.
Return Format
The query returns the start frame or end frame of waveform analysis in integer.