Vital Signs Monitor User Manual Monitoring TEMP
- 96 -
NOTE:
1 The QUICK PREDICT mode is for oral measurement only.
2 Low Temp. Mode can be set to ON only when measure mode is PREDICT.
3 Make sure all settings of TEMP Setup are properly set up every time before you
withdraw the probe from the probe well. If you modify the settings immediately a
measurement is completed, the new settings will be effective for the next
measurement.
14.3 Infrared TEMP with TH Module
14.3.1 Introduction
iM3 with the TH module (Infrared Ear Temperature Module) takes a temperature in the ear.
Diagram of the Infrared Ear Thermometer:
WARNING
1 The infrared ear thermometer is not intended for neonatal patients.
2 Only use the disposable probe covers supplied or recommended by EDAN. Use of
other manufacturer’s probe covers, reuse of disposable probe covers or absence of
probe covers may produce temperature measurement errors and/or inaccuracies.
3 Keep the probe covers away from children.
4 Do not reuse the disposable probe covers.
5 This thermometer converts the ear temperature to display its ‘oral equivalent’
(according to the result of the clinical evaluation to get the offset value). The
thermometer is adjusted to display an oral temperature equivalent. Oral Mode = Ear
Mode + 0.30 °C.