Vital Signs Monitor User Manual Default Settings
- 152 -
C.4 PR Default Settings
PR Settings ADU PED NEO
PR Source SpO
2
Alarm Switch On
Alarm Record Off
Alarm Level Medium
Alarm High Limit 120 160 200
Alarm Low Limit 50 75 100
Pulse Volume 3
Alarm Source SpO
2
or NIBP
C.5 NIBP Default Settings
NIBP Settings ADU PED NEO
Alarm Switch On
Alarm Record Off
Alarm Level Medium
Alarm High Limit (SYS) 160 120 90
Alarm Low Limit (SYS) 90 70 40
Alarm High Limit (MAP) 110 90 70
Alarm Low Limit (MAP) 60 50 30
Alarm High Limit (DIA) 90 70 60
Alarm Low Limit (DIA) 50 40 20
EDAN Module
Inflation value 160 140 100
SunTech Module
Inflation value
160 140 90
Unit mmHg
Interval (Auto) 1 minute
Interval (Average) 1 minute
Measurement times
(Average)
3
PR Switch
On