Vital Signs Monitor User Manual Product Specification
- 132 -
A Product Specification
NOTE:
The performance of the equipment with ☆ mark is determined to be essential
performance.
A.1 Classification
Anti-electroshock type Class I equipment and internal powered equipment
Anti-electroshock degree SpO
2
, NIBP, TEMP: BF
Ingress Protection IPX1
With T2A, TH or F3000 TEMP module: Ordinary
equipment (Sealed equipment without liquid proof)
Disinfection/sterilization method Refer to Chapter Care and Cleaning for details.
Working system Continuous operation equipment
Compliant with Standards IEC 60601-1: 2005+A1 :2012; IEC 60601-1-2: 2014;
EN 60601-1: 2006+A1 :2013; EN 60601-1-2: 2015;
IEC 60601-2-49: 2011;
A.2 Physical Specifications
A.2.1 Size and Weight
Size (159±1) mm (W) × (262±1) mm (H) × (166±1) mm (D)
Weight < 2.5kg (standard configuration, without accessories)
A.2.2 Environment Specification
The monitor may not meet the performance specifications given here if stored or used outside the
specified temperature and humidity ranges.
When the monitor and related products have differing environmental specifications, the effective
range for the combined products is that range which is common to the specifications for all
products.
Temperature
Working
+0 °C to +40 °C (32 ℉~104 ℉)
• With TEMP: +10 °C ~ +40 °C (50 ℉~104 ℉)