Vital Signs Monitor User Manual Alarm Information
- 67 -
TEMP alarm limits are listed as follows:
Patient Type Adjustable Range Step
ADU/PED
+35.5 °C (+95.9 ℉) ~ +42 °C (+107.6 ℉) +0.1 °C (32.2 ℉)
9.5 Known Issues and Directions
No. Known issues Directions
1 When recording the trend graph in
monitor mode, if the Record
shortcut key is pressed to stop
recording in the process of recording
the second parameter trend graph,
the paper for recording will not be
left blank after the second parameter
trend graph.
Re-print trend graph, and do not
press Record shortcut key in
printing process.
2 When USB flash drive is inserted
via OTG interface, it may not be
recognized.
Re-plug USB flash drive
3 In spot-checking mode, press
Admit/Create new patient hardkey
, there is no key sound.
N/A
4 Switching Network Type may lead
to Wi-Fi unavailable.
If Wi-Fi is unavailable, please
restart the monitor (refer to
Section Front View) under the
precondition of ensuring patient’s
safety.
5 In Ward round or spot-checking
mode, the Removable Device can
be selected, while storage device is
fixed as Internal Storage Device.
N/A