Vital Signs Monitor User Manual EMC Information
- 145 -
B EMC Information
- Guidance and Manufacture’s Declaration
B.1 Electromagnetic Emissions
Guidance and manufacture’s declaration – electromagnetic emission
iM3 is intended for use in the electromagnetic environment specified below. The customer or
the user of iM3 should assure that they are used in such an environment.
Emission test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
iM3 uses RF energy only for their internal
function. Therefore, their RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment.
RF emission
CISPR 11
Class A
iM3 is suitable for use in all establishments,
other than domestic establishments and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC/EN 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC/EN 61000-3-3
Complies
NOTE:
The EMISSIONS characteristics of iM3 make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR
11 class B is normally required) iM3 might not offer adequate protection to
radio-frequency communication services. The user might need to take mitigation
measures, such as relocating or re-orienting the equipment.
B.2 Electromagnetic Immunity
Guidance and manufacture’s declaration – electromagnetic immunity
iM3 is intended for use in the electromagnetic environment specified below. The customer or
the user of iM3 should assure that they are used in such an environment.
Immunity test IEC/EN 60601 test level Compliance level
Electromagnetic
environment -
guidance