- 135 -
Each 1 GB extension space for data storage: ≥400 hrs
With all parameters on, storage interval of 1 s, one SpO
2
wave, and one alarm event occurring for
each 10 s.
In Ward round mode, storage data maximally contains the following information:
Ward Round
record
Patient information items (such as MRN,
bed No., patient type, etc.), Ward round data
Up to 800 thousand sets
SpO
2
Storage starting time, SpO
2
value, PR value Up to 20 sets for a single patient
NIBP SYS, DIA, MAP, PR, measurement time Up to 20 sets for a single patient
TEMP TEMP value, measurement time Up to 20 sets for a single patient
Custom
parameters
The custom parameters data Up to 20 sets for a single patient
Each 1 GB space for data storage: ≥100 thousand sets of Ward round records. Up to 800
thousand sets of Ward round records are supported (one Ward round record has 20 original
records).
In spot-checking mode, storage data maximally contains the following information:
16 million sets of spot-checking data for multiple patients.
A.3 NIBP
Complies with IEC 80601-2-30: 2009+A1: 2013.
EDAN Module
Technique Oscillometry
Mode Manual, Auto, Continuous, Average
Measuring Interval in AUTO
Mode (unit: minutes)
1/2/3/4/5/10/15/30/60/90/120/180/240/360/480
Continuous 5 min, interval is 5 s
Measuring Parameter SYS, DIA, MAP, PR
Pressure Unit kPa, mmHg, cmH
2
O
Average measurement Interval (unit: minutes)
1/2/3/4/5
times 3/5
☆Measuring Range