FEASA LED ANALYSER
ICT VERSION
In-Circuit Test Mode
Set the Sensitivity Level of the Analyser (setfactor)
The Sensitivity of the Analyser can be adjusted to allow longer exposure times. This is useful
when testing dim LED's. There are 15 adjustment steps numbered 1 – 15.
Number 1 has the shortest exposure and number 15 will have the longest. The Factory default
value is 1.
Once programmed the Sensitivity Factor remains unchanged even when power is removed.
ONLY RUN THIS TEST DURING INITIAL DEBUG. WHEN SATISIFIED
WITH THE RESULTS COMMENT OUT THIS TEST AS THERE IS A LIMIT
TO THE NUMBER OF TIMES ONE CAN WRITE TO THE ON BOARD FLASH
BEFORE YOU MAY CORRUPT THE DEVICE.