FEASA LED ANALYSER
ICT VERSION
Set Data Mode
set/putLOG - Change the Intensity Response of the
Analyser to Logarithmic mode
Transmit Receive
setlog
putlog
OK
OK
Description
This command will change the Intensity Response of the Analyser to measure in Logarithmic
mode. All Analysers shipped from the factory are set in Log mode.
Please refer to the Intensity Section of this document page 8 Intensity
The set command is stored permanently in memory and can only be changed by using the
setlog command again, power on/off will NOT effect the setlog
The put values are stored in RAM memory and must be changed by using the putlog command
each time after Power On/Off.
Example:
setlog
OK
Example:
putlog
OK
Back to Index