FEASA LED ANALYSER
ICT VERSION
General set/put Commands
setBAUD - Change the baud rate
Transmit Receive
Setbaudxxxxx OK
Where: X = 9600, 19200, 38400, 57600, 115200 for Serial RS232 Port
Description
This command will change the baud rate of the Serial and USB Port's in the Analyser.
The default Port settings of the Analyser are 57,600, 8 Data bits, 1 Stop bit and No Parity.
If an incompatible baud rate is selected the Analyser will not respond. Please disconnect and
reconnect the Analyser to return the Analyser to the default 57600 bps. A correct baud rate
selected will remain stored in memory even after power off.
The command getbaud will return the set baud rate. It can also be seen on the comment
section of the Feasa terminal window.
Example:
To change the baud rate to 19,200 transmit the command setbaud19200 to the Analyser .
setbaud19200
OK