FEASA LED ANALYSER
ICT VERSION
User Calibration Mode
set/putINTGAIN - Set the Intensity GainFactor
Transmit Receive
Setintgain##xxx
putintgain##xxx
OK
OK
Where:
## represents the Fiber Number and is a number in the range 01 – 20.
xxx represents a 3 digit gain factor, default 100.
Description
This command allows the user to adjust the Intensity Gain Factor for each Fiber.
This is useful when it is required to balance all or some of the Fibers to give the same Intensity
when testing similar LED's. The Factory default setting is 100 and the value for each Fiber
can be adjusted from 050 – 200.
These set values are stored permanently in memory and can only be changed by using the
setIntGain command again.
The put values are stored in RAM memory and must be changed by using the putIntGain
command each time after Power On/Off.
The command getIntGain will display the current stored gain setting.
Example:
Set the Intensity gain for Fiber 1 to 095.
setintgain01095
OK
Example:
Set the Intensity gain for Fiber 1 to 095.
putintgain01095
OK
Back to Index
89

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Feasa ICT Series and is the answer not in the manual?

Feasa ICT Series Specifications

General IconGeneral
BrandFeasa
ModelICT Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals