Specifications 10-5
August 2003
Dew/frost Point Temperature
Overall Calibration
Range: -110°C to 60°C
Calibration Range
Options:
Standard: 20° to -80°C with data to -110°C
Ultralow: -50 to -110°C
Extended High: 60° to -80°C with data to -110°C
Accuracy: ±2°C from 60° to -65°C
±3°C from -65° to -110°C
Repeatability: ±0.5°C from 60° to -65°C
±1°C from -65° to -110°C
Operating Pressure: 5 µ of Hg to 5,000 psig (limited by
pressure sensor — see pressure
measurement specifications)
Temperature Measurement
Note: An optional thermistor is available for all GE
Panametrics moisture probes.
Range: -30°C to 70°C
Accuracy: ±0.5°C at -30°C