EasyManuals Logo

GE Venue Go Service Manual

GE Venue Go
324 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #239 background imageLoading...
Page #239 background image
D
IRECTION 5813707-100, REVISION 2 VENUE™ SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-33
PRELIMINARY
8-2-12 Speakers Replacement Procedure
8-2-12-1 Tools
Phillips screwdriver of appropriate size
8-2-12-2 Time Required
5 min
8-2-12-3 Preparations
1.) Shut down the Venue Go™ ultrasound unit, as described in Power Shutdown.
2.) Prepare stable flat surface covered with a soft material (such as foam or thick cloth) to prevent
scratches to the display.
3.) Dismount Venue Go™ system from cart as described in Mounting / Dismounting System on Cart
and place it on a previously prepared surface with the display facing down.
8-2-12-4 Speakers Removal Procedure
1) Remove back cover.
2) Remove the fan.
3.) Release the speakers cables from the cable clips.
FRU Part #
Refer to Table 9-16 on page 9-17.
Back Cover Removal Procedure
Fan Removal Procedure
Figure 8-109 Speakers Cable