IRECTION 5813707-100, REVISION 2 VENUE™ SERVICE MANUAL
8-46 Chapter 8 - Replacement Procedures
PRELIMINARY
8-2-18 Printer Assembly Replacement Procedure
8-2-18-1 Tools
Phillips screwdriver of appropriate size
8-2-18-2 Time Required
5 min
8-2-18-3 Preparations
1.) Shut down the Venue Go™ ultrasound unit, as described in Power Shutdown .
2.) Prepare stable flat surface covered with a soft material (such as foam) to prevent scratches to the
display.
3.) Dismount Venue Go™ system from cart and place it on a previously prepared surface with the
display facing down.
8-2-18-4
Printer Assembly Removal Procedure
4.) Disconnect printer cables: USB, printer power supply.
FRU Part #
Refer to Table 9-16 on page 9-17.
Disconnect printer PWR supply cable
Figure 8-126 Disconnect printer USB and Power Cables
Disconnect
printer USB
Disconnect
printer PWR