EasyManuals Logo

GE Venue Go Service Manual

GE Venue Go
324 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #254 background imageLoading...
Page #254 background image
D
IRECTION 5813707-100, REVISION 2 VENUE™ SERVICE MANUAL
8-48 Chapter 8 - Replacement Procedures
PRELIMINARY
8-2-19 Printer and Printer Bracket Replacement Procedure
8-2-19-1 Tools
Phillips screwdriver of appropriate size
8-2-19-2 Time Required
5 min
8-2-19-3 Preparations
1.) Shut down the Venue Go™ ultrasound unit, as described in Power Shutdown .
2.) Prepare stable flat surface covered with a soft material (such as foam) to prevent scratches to the
display.
3.) Dismount Venue Go™ system from cart and place it on a previously prepared surface with the
display facing down.
8-2-19-4 Printer and Printer Bracket Removal Procedure
4.) Remove printer assembly
5.) Remove 4 Phillips screws holding the printer inside the printer tray.
FRU Part #
Refer to Table 9-16 on page 9-17.
Printer Assembly Removal Procedure
Figure 8-128 Disconnect Printer Screws