D
IRECTION 5813707-100, REVISION 2 VENUE™ SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-71
PRELIMINARY
8-3-4 Software Reload/Update Procedure
NOTE: Images in this procedure are for reference only. There may be differences or variations, in accordance
with different software versions.
8-3-4-1 Backup System Configuration
Perform backup of all system configuration parameters as follows:
1.) Log in to the system as ADM (administrator).
2.) Tap Settings and select Config.
3.) On the Connectivity tab, select Dataflow and make sure the USB Storage option is enabled.
4.) Connect USB memstick to any available USB port in the system.
Figure 8-157 Logging In as ADM
Figure 8-158 Backup Options Tab